15B MAX/17B MAX
Users Manual
16
Specifications
General Specifications
Maximum voltage between
any Terminal and Earth Ground..............600 V
Maximum differential voltage
between V and COM terminals...............1000 V
Display (LCD)..................................................6000 counts, updates
3/sec
Battery Type...................................................2 AA, IEC LR6
Battery Life.....................................................500 hours minimum
Te m p e r a t u r e
Operating.....................................................0 °C to 40 °C
Storage.........................................................-30 °C to 60 °C
Relative Humidity
Operating Humidity .................................Non-condensing
(<10 °C); ≤90 % RH at
10 °C to 30 °C;
≤75 % RH at 30 °C to
40 °C
Operating Humidity, 40 MΩ range.... ≤80 % RH at 10 °C to
30 °C; ≤70 % RH at
30 °C to 40 °C
Altitude
Operating..................................................2000 m
Storage......................................................12 000 m
Temperature Coefficient...........................0.1 X (specified
accuracy) /°C (<18 °C or
>28 °C)
WFuse protection for
current inputs................................................0.44 A, 1000 V, IR 10 kA
11 A, 1000 V, IR 20 kA
Size (HxWxL) ..................................................183 mm x 91 mm x
49.5 mm
Weight...............................................................455 g
Ingress Protection.......................................IEC 60529: IP40
Safety
General .......................................................IEC 61010-1: Pollution
Degree 2
Measurement ...........................................IEC 61010-2-033:
CAT III 600V
Electromagnetic Environment ..............IEC 61326-1: Portable