15B MAX/17B MAX
Users Manual
18
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at
operating temperatures of 18 °C to 28 °C, relative humidity
at 0 % to 75 %. Accuracy specifications take the form of:
±([% of Reading] + [Number of Least Significant Digits]).
AC and DC Voltage
Function Range Resolution
Accuracy
15B MAX 17B MAX
AC Volts (40 Hz to 500 Hz)
[1]
K
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1V
1.0 % + 3 1.0 % + 3
AC Millivolts
u
600.0 mV 0.1 mV 3.0 % + 3 3.0 % + 3
DC Millivolts
M
600.0 mV 0.1 mV 1.0 % + 10 1.0 % + 10
DC Volts
V
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1V
0.5 % + 3 0.5 % + 3
[1] All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.