Fluke 19xC-2x5
Users Manual
94
Figure 54. Parasitic capacitance between probes,
instrument and environment
M
MOTION/MOTOR CONTROLLER
DC BUS
DC BUS
DIGITAL
GROUND
-
Figure 55. Parasitic capacitance between analog and
digital reference
NAL
INPUT DI
ITAL
NTR
LLER
ANALOG
GROUND
DIGITAL
GROUND
Figure 56. Correct connection of reference leads
NAL
INPUT DI
ITAL
NTR
LLER
ANALOG
GROUND
DIGITAL
GROUND
Figure 57. Wrong connection of reference leads
Noise that is picked up by reference lead B can be
transmitted by parasitic capacitance to the analog input
amplifier.