Fluke 43
Users Manual
32
Safety Specifications
Safety Characteristics
Designed and tested for measurements on 600 V rms Category III,
Pollution Degree 2 in accordance with:
EN 61010.1 (1993) (IEC 1010-1)
ANSI/ISA S82.01-1994
CAN/CSA-C22.2 No.1010.1-92 (including approval)
UL3111-1 (including approval)
Installation Category III refers to distribution level and fixed installation
circuits inside a building.
Maximum input voltage Input 1 and 2
Direct on inputs or with test leads TL24 (see Figure 11)
0 to 66 kHz .................................................................................600 V rms
> 66 kHz .......................................................................derating to 5 V rms
With Shielded Banana-to-BNC Adapter Plug BB120 (see Figure 11)
0 to 400 kHz ...............................................................................300 V rms
> 400 kHz .....................................................................derating to 5 V rms
Figure 11. Max. Input Voltage v.s. Frequency
Maximum floating voltage
From any terminal to ground
0 to 400 Hz .................................................................................600 V rms