Specifications
Function Specifications
3
37
Meter
Ohm
Ranges ..........................................................500.0
Ω
to 5.000 M
Ω
, 30.00 M
Ω
±(0.6 % + 5 counts)
Max. Measurement Current..................................................................0.5 mA
Measurement Voltage at open circuit...................................................... < 4V
Diode
Accuracy...............................................................................±(2 % +5 counts)
Max. Measurement Current..................................................................0.5 mA
Measurement Voltage at open circuit..................................................... < 4 V
Continuity
Beep .........................................................................................< 30
Ω
(± 5
Ω
)
Measurement Current...........................................................................0.5 mA
Detection of shorts................................................................................> 1 ms
Capacitance
Ranges ...........................................................................50.00 nF to 500.0
µ
F
±(2 % +10 counts)
Max. Measurement Current..................................................................0.5 mA
Temperature
Ranges (°C or °F).........................-100.0 to +400.0 °C or -200.0 to +800.0 °F
±(0.5 % + 5 counts)
Record
Record times (selectable).............................. 4 min to 8 days, endless (16 days)
Number of readings............................................................1 or 2 simultaneously
Record is available for the functions:
•
volts / amps / hertz
•
power
•
harmonics
•
ohms / continuity / capacitance
•
temperature
•
scope