5725A
Instruction Manual
1-24
Table 1-17. 5720A Series II AC Current Specifications: 95 % Confidence Level
5720A
95%
95 % Confidence Level
Range Resolution Frequency
Absolute Uncertainty
±5 °C from calibration temperature
For fields strengths >0.4 V/m but 3 V/m,
add 1 % of range.
Relative Uncertainty
±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Hz ± (ppm output + nA) ± (ppm output + nA)
220 µA1 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
05k - 10 k
0210 + 16
0130 + 10
0100 + 8
0240 + 12
0800 + 65
0230 + 16
0140 + 10
0110 + 8
0250 + 12
0900 + 65
0240 + 16
0150 + 10
0115 + 8
0270 + 12
1000 + 65
0250 + 16
0160 + 10
0120 + 8
0280 + 12
1100 + 65
0210 + 16
0110 + 10
0080 + 8
0200 + 12
0700 + 65
0230 + 16
0130 + 10
0090 + 8
0230 + 12
0800 + 65
2.2 mA 10 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
05k - 10 k
0210 + 40
0140 + 35
0100 + 35
0170 + 110
0800 + 650
0230 + 40
0140 + 35
0110 + 35
0180 + 110
0900 + 650
0240 + 40
0150 + 35
0115 + 35
0190 + 110
1000 + 650
0250 + 40
0160 + 35
0120 + 35
0200 + 110
1100 + 650
0210 + 40
0110 + 35
0080 + 35
0200 + 110
0700 + 650
0230 + 40
0130 + 35
0090 + 35
0230 + 110
0800 + 650
22 mA 100 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
05k - 10 k
0210 + 400
0130 + 350
0100 + 350
0170 + 550
0800 + 5000
0230 + 400
0140 + 350
0110 + 350
0180 + 550
0900 + 5000
0240 + 400
0150 + 350
0115 + 350
0190 + 550
1000 + 5000
0250 + 400
0160 + 350
0120 + 350
0200 + 550
1100 + 5000
0210 + 400
0110 + 350
0080 + 350
0200 + 550
0700 + 5000
0230 + 400
0130 + 350
0090 + 350
0230 + 550
0800 + 5000
Hz ± (ppm output + µA) ± (ppm output + µA)
220 mA 1 µA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
05k - 10 k
0210 + 4
0130 + 3.5
0100 + 2.5
0170 + 3.5
0800 + 10
0230 + 4
0140 + 3.5
0110 + 2.5
0180 + 3.5
0900 + 10
0240 + 4
0150 + 3.5
0115 + 2.5
0190 + 3.5
1000 + 10
0250 + 4
0160 + 3.5
0120 + 2.5
0200 + 3.5
1100 + 10
0210 + 4
0110 + 3.5
0080 + 2.5
0200 + 3.5
0700 + 10
0230 + 4
0130 + 3.5
090 + 2.5
0230 + 3.5
0800 + 10
2.2 A 10 µA
020 - 1 k
01 k - 5 k
05 k - 10 k
0230 + 35
0350 + 80
5000 + 160
0240 + 35
0390 + 80
6000 + 160
0250 + 35
0420 + 80
6500 + 160
0260 + 35
0450 + 80
7000 + 160
0250 + 35
0400 + 80
5000 + 160
0300 + 35
0440 + 80
6000 + 160
5725A Amplifier:
11 A 100 µA
040 - 1 k
01 k - 5 k
05 k - 10 k
0370 + 170
0800 + 380
3000 + 750
0400 + 170
0850 + 380
3300 + 750
0440 + 170
0900 + 380
3500 + 750
0460 + 170
0950 + 380
3600 + 750
0300 + 170
0700 + 380
2800 + 750
0330 + 170
0800 + 380
3200 + 750
Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 µA and
2.2 mA ranges are increased by 1.3 x plus 2 µA when supplied through 5725A terminals. Specifications are
otherwise identical for all output locations.