Introduction and Specifications
AC Current Specifications
1
1-25
Table 1-18. 5700A Series II AC Current Specifications: 99 % Confidence Level
5700A
99%
99 % Confidence Level
Range Resolution Frequency
Absolute Uncertainty
±5 °C from calibration temperature
For fields strengths >0.4 V/m but 3 V/m,
add 1 % of range.
Relative Uncertainty
±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Hz ± (ppm output + nA) ± (ppm output + nA)
220 µA1 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
0
5k - 10 k
0650 + 30
0350 + 25
0120 + 20
0500 + 50
1500 + 100
0700 + 30
0380 + 25
0140 + 20
0600 + 50
1600 + 100
0750 + 30
0410 + 25
0150 + 20
0650 + 50
1700 + 100
00 800 + 30
00 420 + 25
00 160 + 20
00 700 + 50
0 1800 + 100
0450 + 30
0270 + 25
0110 + 20
0450 + 50
1400 + 100
0500 + 30
0300 + 25
0120 + 20
0500 + 50
1500 + 100
2.2 mA 10 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
0
5k - 10 k
0650 + 50
0350 + 40
0120 + 40
0500 + 500
1500 + 1000
0700 + 50
0380 + 40
0140 + 40
0600 + 500
1600 + 1000
0750 + 50
0410 + 40
0150 + 40
0650 + 500
1700 + 1000
00 800 + 50
0 0420 + 40
00 160 + 40
00 700 + 500
0 1800 + 1000
0450 + 50
0270 + 40
0110 + 40
0450 + 500
1400 + 1000
0500 + 50
0300 + 40
0120 + 40
0500 + 500
1500 + 1000
22 mA 100 nA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
0
5k - 10 k
0650 + 500
0350 + 400
0120 + 400
0500 + 5000
1500 + 10,000
0700 + 500
0380 + 400
0140 + 400
0600 + 5000
1600 + 10,000
0750 + 500
0410 + 400
0150 + 400
0650 + 5000
1700 + 10,000
00 800 + 500
00 420 + 400
00 160 + 400
00 700 + 5000
0 1800 + 10,000
0450 + 500
0270 + 400
0110 + 400
0450 + 5000
1400 + 10,000
0500 + 500
0300 + 400
0120 + 400
0500 + 5000
1500 +
10,000
Hz ± (ppm output + µA) ± (ppm output + µA)
220 mA 1 µA
010 - 20
020 - 40
040 - 1 k
01k - 5 k
0
5k - 10 k
0650 + 5
0350 + 4
0120 + 4
0500 + 50
1500 + 100
0700 + 5
0380 + 4
0150 + 4
0600 + 50
1600 + 100
0750 + 5
0410 + 4
0170 + 4
0650 + 50
1700 + 100
00 800 + 5
00 420 + 4
00 180 + 4
0 700 + 50
0 1800 + 100
0450 + 5
0280 + 4
0110 + 4
0450 + 50
1400 + 100
0500 + 5
0300 + 4
0130 + 4
0500 + 50
1500 + 100
2.2 A 10 µA
0
20 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0600 + 40
0700 + 100
8000 + 200
0650 + 40
0750 + 100
9000 + 200
0700 + 40
0800 + 100
9500 + 200
0 750 + 40
00 850 + 100
10,000 + 200
0600 + 40
0650 + 100
7500 + 200
0650 + 40
0750 + 100
8500 + 200
5725A Amplifier:
11 A 100 µA
0
40 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0370 + 170
0800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
0440 + 170
0900 + 380
3500 + 750
00 460 + 170
00 950 + 380
0 3600 + 750
0300 + 170
0700 + 380
2800 + 750
0330 + 170
0800 + 380
3200 + 750
Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 µA and 2.2 mA
ranges are increased by 1.3 x plus 2 µA when supplied through 5725A terminals. Specifications are otherwise
identical for all output locations.