Appendices
Calibration Constant Information C
C-11
Group DC_22V
Name Default Function
DI_22V 0.5 22V range DAC volts per input volt
OF_22V 0.0 22V range DC offset
Group ZC_22V
Name Default Function
Z_22V 0.0 22V range zero
SHO_22V 0.0 22V AUX input offset
IA_22V 10.0 22V range A/D volts per input volt
Group AC_22V
Name Default Function
F1_22V 1.0 22V range flatness correction 1 (10 Hz)
F2_22V 1.0 22V range flatness correction 2 (100 Hz)
F3_22V 1.0 22V range flatness correction 3 (1 kHz)
F4_22V 1.0 22V range flatness correction 4 (10 kHz)
F5_22V 1.0 22V range flatness correction 5 (20 kHz)
F6_22V 1.0 22V range flatness correction 6 (50 kHz)
F7_22V 1.0 22V range flatness correction 7 (100 kHz)
Group ZC_22VHF
Name Default Function
Z_22VHF 0.0 22VHF range zero
SHO_22VHF 0.0 22VHF AUX input offset
IA_22VHF 10.0 22VHF range A/D volts per input volt
Group AC_22VHF
Name Default Function
F1_22VHF 1.0 22VHF range flatness correction 1 (100 kHz)
F2_22VHF 1.0 22VHF range flatness correction 2 (300 kHz)
F3_22VHF 1.0 22VHF range flatness correction 3 (500 kHz)
F4_22VHF 1.0 22VHF range flatness correction 4 (800 kHz)
F5_22VHF 1.0 22VHF range flatness correction 5 (1 MHz)