6109A/7109A
Operators Manual
52
INITiate
(-P models only) Measurement is continuous and automatic, so no triggering is
necessary.
INITiate:CONTinuous 1
INITiate:CONTinuous?
(-P models only) Measurement is always continuous.
INPut2:MAMP:LPOWer <boolean>|DEFault
INPut2:MAMP:LPOWer?
(-P models only) Turn transmitter loop power on or off.
<boolean> is 0 (OFF, default) or 1(ON).
This command is conditionally password protected.
INPut2:RTD:WIRes 2|3|4|DEFault
INPut2:RTD:WIRes?
(-P models only) Select the number of lead wires for the DUT RTD circuit. The
default is 4 wires.
This command is conditionally password protected.
MEASure?
(-P models only) Return the reference probe temperature and DUT sensor
reading.
The response is temperature in °C or °F and depends on the temperature unit
setting.
OUTPut:DATA?
Return percent heating output of the temperature control function.
The response is a decimal number of percent. The value is negative as the
Product cools. The value is 0 if control is off.
OUTPut:STATe <boolean>|DEFault
OUTPut:STATe?
Turn temperature control on or off.
<boolean> is 0 (OFF, default) or 1(ON).
*RST sets this to off.
Example:
OUTP:STATE 1