Portable Calibration Baths
Commands
53
PROGram:BEEP <boolean>|DEFault
PROGram:BEEP?
Turn program beep on or off.
<boolean> is 0 (OFF) or 1(ON, default).
This command is conditionally password protected.
PROGram:CATalog?
Return a list of defined programs.
The response is a series of comma-separated program names, each enclosed in
quotation marks.
PROGram:DELete:ALL
Delete all defined programs.
Because at least one program must be defined, a new default program will be
created to replace those that have been deleted.
This command is password protected.
PROGram:EXPlicit:DEFine <name>[,<numeric_value>[,...]]
PROGram:EXPlicit:DEFine?
Create a new temperature program.
<name> is the name of the new program enclosed in quotation marks. It can
contain up to 14 characters. Characters can range from ‘A’ to ‘Z’, ‘a’ to ‘z’, ‘0’ to
‘9’, ‘-‘, ‘.’, ‘/’, and ‘_’. The name must be different from any other defined program.
<numeric_value> are values of program variables. If parameters are omitted,
default values are assumed. Variable are defined the order shown in Table 7.