Users Guide
7-6
1. This is the top half of the latch which attaches the lid to the bath. There are
two latches one on the front and one on the back.
2. The thermometer/liquid level is a hole in the lid for filling and removing of
liquid, checking liquid level and for mounting a thermometer. A hole may be
drilled in the rubber stopper supplied for inserting a thermometer into the
bath.
3. The rectangular access hole may be used to insert the user’s device to be
thermo stated. To obtain specified control levels this hole must be covered to
prevent thermal disturbance to the bath. Covers may be purchased
separately from Hart Scientific.
4. The circulation inlet - outlet is the access area or the inlet and outlet for the
circulation pump.
5. The stirrer/pump motor is mounted on the bath tank lid. It drives the stirring
propellers to provide mixing of the bath fluid. Proper mixing of the fluid is
important for good constant temperature stability and uniformity. The
stirrer/pump motor power cord plugs into the back at the power socket
labeled “STIRRER”. The pump motor (on optional pump lid) is mounted on
the bath tank lid.
6. The probe hole is for the temperature controller probe. The probe must be in
this hole for the bath to operate.