Calibration Procedure
11.2 Measuring the Set-point Error11
11-3
D0 = 25.229
DG = 0.0028530
t
L
= 75.00 C
measured t =74.901 C
Compute errors,
err
L
= 24.869 25.00 °C = 0.131 °C
err
H
= 74.901 75.00 °C = 0.099 °C
Compute C
0
,
392.25)229.25(
0.250.75
))229.25(0.25)(099.0())229.25(0.75)(131.0(
0 =+
D
Compute C
G
,
0028548.00028530.01
0.250.75
)131.0()099.0(
=
+
=
GD
Figure 10. Sample calibration computations