RTD/Thermocouple Calibrator
General Specifications
Maximum voltage applied between any
terminal and earth ground or between any
two terminals:
30 V
Operating temperature
-10 °C to 50 °C
Storage temperature
-20 °C to 60 °C
Operating altitude 2,000 meters
Storage altitude 12,000 meters
Relative Humidity (% RH operating without
condensation)
Non condensing
90 % (10 °C to 30 °C)
75 % (30 °C to 40 °C)
45 % (40 °C to 50 °C)
(Without condensation)
Vibration Requirements MIL-T-28800E, Class 2
Drop Test Requirements 1 meter
IP Rating IEC 60529: IP52
Electromagnetic Environment
IEC 61326-1, Portable
FCC: CFR Title 47, Part 15, Subpart B
Safety IEC 61010-1, Max 30 V to earth, Pollution Degree 2
Electromagnetic Compatibility
Applies to use in Korea only. Class A Equipment (Industrial
Broadcasting & Communication Equipment)
This product meets requirements for industrial (Class A)
electromagnetic wave equipment and the seller or user should
take notice of it. This equipment is intended for use in business
environments and is not to be used in homes.
Power Supply 4 AA alkaline batteries/NEDA code: 15A, IEC code: LR6
Size (H x W x L) 52.5 x 84 x 188.5 mm
Weight 524 g
5

Table of Contents

Other manuals for Fluke 712B/EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 712B/EN and is the answer not in the manual?

Fluke 712B/EN Specifications

General IconGeneral
BrandFluke
Model712B/EN
CategoryTest Equipment
LanguageEnglish

Related product manuals