714B Calibration Adjustment
Save Data and Reset the 712B System
After the calibration procedure described above is done, push to confirm and
store the calibration data. The 712B system is automatically reset.
hwp18.jpg hwp19.jpg
714B Calibration Adjustment
Enter Calibration Mode
To enter manual calibration mode:
1. Power off the Product.
2. Push and hold down and at the same time.
3. Push and then release .
4. Release and .
The Welcome screen shows.
5. After a warmup period of approximately 5 minutes, start the calibration
adjustment.