RTD/Thermocouple Calibrator
714B Calibration Adjustment
Fluke 9101-256
Fluke 714B
J TYPE TC Sensor
hwp008.eps
Figure 17. Connect the 714B to the Fluke 9101-256
2. After the J type thermocouple is inserted, wait approximately 5 minutes for
the temperature reading to stabilize.
hwp25.jpg
3. Push to calibrate the cold junction temperature.
35