772/773
Calibration Manual
6
Specifications
Electrical Specifications
DC Current Measurement
With Jaw
Ranges ........................................... 0 mA to 20.99 mA, 21 mA to 100 mA
Resolution ................................. 0.01 mA, 0.1 mA
Accuracy .................................... 0.2 % + 5 counts, 1 % + 5 counts
In Circuit
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
Current Source
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
mA Drive .................................... 24 mA into 1000 Ω
Current Simulate
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
Maximum Voltage ...................... 50 V
DC Voltage Measurement (773)
Range ........................................ 0-30 V
Resolution .................................. 0.01 V
Accuracy .................................... 0.2 % + 2 counts
DC Voltage Source (773)
Range ........................................ 0 V to 30 V
Resolution ................................. 0.01 V
Accuracy .................................... 0.2 % + 2 counts
mA Drive .................................... 2 mA maximum all conditions
mA DC IN/OUT (773)
Sourcing range .......................... 0 mA to 24 mA
Sourcing resolution .................... 0.01 mA
Sourcing accuracy ..................... 0.2 % + 2 counts
Measurement range .................. 0 mA to 24 mA
Measurement resolution ............ 0.01 mA
Measurement accuracy ............. 1 % FS
Scaled mA DC Current Output to mA Current Input from the Jaw (773)
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 1 % FS
Response speed ........................ 2x/second
DC Loop Power ......................... 24 V
Influence of Earth’s Field ........... <0.20 mA
Batteries .................................... 4 1.5 V, Alkaline, IEC LR6
Working hours ........................... 12 hours @ 12 mA sourced into 500 Ω