EasyManuals Logo

Fluke 773 User Manual

Fluke 773
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
772/773
Calibration Manual
6
Specifications
Electrical Specifications
DC Current Measurement
With Jaw
Ranges ........................................... 0 mA to 20.99 mA, 21 mA to 100 mA
Resolution ................................. 0.01 mA, 0.1 mA
Accuracy .................................... 0.2 % + 5 counts, 1 % + 5 counts
In Circuit
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
Current Source
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
mA Drive .................................... 24 mA into 1000 Ω
Current Simulate
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 0.2 % + 2 counts
Maximum Voltage ...................... 50 V
DC Voltage Measurement (773)
Range ........................................ 0-30 V
Resolution .................................. 0.01 V
Accuracy .................................... 0.2 % + 2 counts
DC Voltage Source (773)
Range ........................................ 0 V to 30 V
Resolution ................................. 0.01 V
Accuracy .................................... 0.2 % + 2 counts
mA Drive .................................... 2 mA maximum all conditions
mA DC IN/OUT (773)
Sourcing range .......................... 0 mA to 24 mA
Sourcing resolution .................... 0.01 mA
Sourcing accuracy ..................... 0.2 % + 2 counts
Measurement range .................. 0 mA to 24 mA
Measurement resolution ............ 0.01 mA
Measurement accuracy ............. 1 % FS
Scaled mA DC Current Output to mA Current Input from the Jaw (773)
Range ........................................ 0 mA to 24 mA
Resolution ................................. 0.01 mA
Accuracy .................................... 1 % FS
Response speed ........................ 2x/second
DC Loop Power ......................... 24 V
Influence of Earth’s Field ........... <0.20 mA
Batteries .................................... 4 1.5 V, Alkaline, IEC LR6
Working hours ........................... 12 hours @ 12 mA sourced into 500 Ω

Other manuals for Fluke 773

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 773 and is the answer not in the manual?

Fluke 773 Specifications

General IconGeneral
Operating altitude0 - 2000 m
Storage temperature (T-T)-25 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)0 - 90 %
Battery typeAA
Power sourceBattery
Battery voltage1.5 V
Battery technologyAlkaline
Number of batteries supported4
Product typeDigital multimeter
Product colorGrey, Red, Yellow
DC current range0 - 0.024 A
DC voltage range0 - 30 V
Placement supportedHandheld multimeter
Basic accuracy (DC current)0.2% + 2 counts
Weight and Dimensions IconWeight and Dimensions
Depth44 mm
Width70 mm
Height246 mm
Weight410 g

Related product manuals