Milliamp Process Clamp Meter
Accuracy Tests
13
Volts DC Measure Accuracy Tests
Step Calibrator Output
UUT Meter Reading Limit
Low High
1 0.00 V -0.02 V 0.02 V
2 10.00 V 9.96 V 10.04 V
3 -10.00 V -10.04 V -9.96 V
4 20.00 V 19.94 V 20.06 V
5 -20.00 V -20.06 V -19.94 V
6 30.00 V 29.92 V 30.08 V
7 -30.00 V -30.08 V -29.92 V
mA DC Source Accuracy Tests
Step UUT Meter Output
DMM Reading Limit
Low High
1 0.00 mA -0.02 mA 0.02 mA
2 4.00 mA 3.97 mA 4.03 mA
3 8.00 mA 7.96 mA 8.04 mA
4 12.00 mA 11.96 mA 12.04 mA
5 20.00 mA 19.94 mA 20.06 mA
6 24.00 mA 23.93 mA 24.07 mA
Volts DC Source Accuracy Tests
Step UUT Meter Output
DMM Reading Limit
Low High
1 0.00 V -0.02 V 0.02 V
2 2.50 V 2.47 V 2.53 V
3 5.00 V 4.97 V 5.03 V
4 7.50 V 7.46 V 7.54 V
5 10.00 V 9.96 V 10.04 V