1
Introduction
XW Warning
To prevent electrical shock or personal injury, do not do the
calibration verification tests or calibration procedures in this
manual unless you are qualified.
The data in this manual is for qualified personnel only.
This manual tells you about verification and adjustment procedures for the
772/773 Milliamp Process Clamp Meter (referred to in this manual as the Meter
or Product). The Meter features closed-case calibration to use with reference
sources. It measures the reference signals, calculates the correction factors, and
keeps them in memory. Calibration adjustment is required after a repair, or if the
Meter fails a performance test.
This manual explains:
• Precautions and Safety Information
• Specifications
• Basic Maintenance
• Calibration/Verification Procedure
• Replaceable Parts and Accessories
For complete use instructions, refer to the 772/773 Instruction Sheet.