Reference Multimeter and 8 ½ Digit Multimeter
Instrument Setup
99
Instrument Settings
Use the Instrument Settings menu to change global instrument setting
parameters. Push (Instrument Settings) to view the Instrument Settings
submenu. See Table 22.
Table 22. Instrument Settings Parameters
Menu Parameter Change the Parameter
Line Frequency
The Product automatically detects the mains frequency but a specific line
frequency can be set here in non-volatile memory. To change the Line
Frequency, use the navigation keys to move the cursor to highlight this field
and then push (Auto) or (50 Hz) (60 Hz). The Auto setting
measures the line frequency at the time the Auto settings is made and at
Product power-on. It does not continually monitor the line frequency nor does it
use an off-nominal value.
Ext. Ref. Clk
(External Reference
Clock)
When on, this enables the rear-panel FREQ REF IN input to use an external
1 MHz or 10 MHz clock. To change the External Reference Clock, use the
navigation keys to move the cursor to highlight this field and then push
(OFF) or (1MHz) or (10MHz).

Table of Contents

Other manuals for Fluke 8558A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8558A and is the answer not in the manual?

Fluke 8558A Specifications

General IconGeneral
BrandFluke
Model8558A
CategoryMultimeter
LanguageEnglish

Related product manuals