Reference Multimeter and 8 ½ Digit Multimeter
Mains Voltage
7
Mains Voltage
XW Warning
To prevent possible electrical shock, fire, or personal injury:
• Do not put the Product where access to the mains power cord
is blocked.
• Use onl
y the mains power cord a
nd connector approved for
the voltage and plug configuration in
your country and
rated
for the Product.
• Make sure the ground conductor in the mains po
w
er cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.
• Replace the mains pow
er cord if t
he insulation is damaged or
if the insulation shows signs of wear.
• The Product enclosure must be grounded through the
grounding conductor of the po
we
r cord, or through the rear
panel ground binding post.
The Product comes with the appropriate line power plug for the country of
purchase. If a different type is necessary, see Table 2. They list and show the
mains line power plug types available from Fluke Calibration.
The Product automatically detects the main line voltage when powered up and
configures itself to work at that voltage level. Nominal mains voltages ranging
from 100 V rms to 120 V rms and from 220 V rms to 240 V rms ( each ±10 %) are
acceptable, with frequencies from 47 Hz to 63 Hz.