Reference Multimeter and 8 ½ Digit Multimeter
Functions
21
AC Voltage Menu
This section explains the AC Voltage (ACV) menu. See the screen below.
iei005.png
(Range): Select each of the ac V ranges manually or select Auto to put
the Product into Autorange. Make the range selection with the softkeys or use
the navigational keys to highlight the selection and push .
(Resolution): ACV has resolution from 4 1/2 digits to 7 1/2 digits. The
default is 6 1/2 digits. To choose the resolution, use the softkeys or the
navigational keys. Push the navigational keys to highlight the selection and then
push .
(Band): ACV has selectable-bandwidth settings.
The Product has these available settings:
Wideband (default)
Extended High Frequency
Most applications should use Wideband which measures signals up to 2 MHz
and where the wave shape of the signal is not necessarily known. It is the default
setting and is a general-purpose ac voltage measurement function.
Extended High Frequency extends the ACV frequency range to 10 MHz. This
mode is about three times slower and less accurate than Wideband.