Reference Multimeter and 8 ½ Digit Multimeter
Functions
35
Current stimulus values are shown in Table 9 for each of the five resistance
modes.
Table 9. Ohms Stimulus Levels for Each Mode
Range
2W and 4W
Normal
2W and 4W
Normal
with LoI
ON
4W Tru Ω
Tru Ω LoI
4W Tru
ohm with
LoI ON
2W and 4W
HV Ω
1 Ω 100 mA 100 mA ±100 mA ±100 mA NA
10 Ω 10 mA 10 mA ±10 mA ±10 mA NA
100 Ω 10 mA 1 mA ±10 mA ±1 mA NA
1 kΩ 1 mA 100 µA ±1 mA ±100 µA NA
10 kΩ 100 µA 10 µA ±100 µA ±10 µA NA
100 kΩ 100 µA 10 µA NA NA NA
1M Ω 10 µA 1 µA NA NA NA
10 MΩ 1 µA 100 nA NA NA 10 µA
100 MΩ 100 nA 10 nA NA NA 1 µA
1 GΩ 10 nA 10 nA NA NA 100 nA
10 GΩ NA NA NA NA 10 nA