8588A/8558A
Operators Manual
70
Note
4-wire PRT uses True Ohms so zeroing is not required.
Connect the PRT probe to the Product in the same way as when making
resistance measurements, using the appropriate connection shown in Figure 19.
Select the corresponding 2-, 3-, or 4-wire probe type using the (Probe)
softkey. Fluke Calibration recommends that Ext. Guard is ON (,
(Ext. Guard)).
adj131f.emf
Figure 19. PRT Connections
Note
The 3-wire PRT connection is actually a 4-wire measurement, and
requires a short between the low terminals as depicted in Figure 19.