True-rms Multimeter
Detailed Specifications
49
Current Function
Function Range Resolution Accuracy
[1, 2]
Burden Voltage (typical)
mA
\
(45 Hz to 2 kHz)
mA
[
60.00 mA
400.0 mA
[4]
6.000 A
10.00 A
[3]
60.00 mA
400.0 mA
[4]
6.000 A
10.00 A
[3]
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 mA
0.1 mA
0.001 A
0.01 A
±(1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
µA B
(45 Hz to 2 kHz)
µAF
600.0 µA
6000 µA
600.0 µA
6000 µA
0.1 µA
1 µA
0.1 µA
1 µA
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
100 µV/µA
100 µV/µA
100 µV/µA
100 µV/µA
[1] AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range (5 % to 100 % of range) and 10 A
range (15 % to 100 % or range).
[2] The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading
between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of range. Using REL to offset this reading
may produce a much larger constant error in later measurements
[3] W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A unspecified.
[4] 400 mA continuous; 600 mA for 18 hrs maximum.