EasyManua.ls Logo

Fluke 87V Ex - Using Conductance for High Resistance or Leakage Tests

Fluke 87V Ex
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87V Ex
Users Manual
22
Using Conductance for High Resistance or
Leakage Tests
Conductance, the inverse of resistance, is the ability of a
circuit to pass current. High values of conductance
correspond to low values of resistance.
The Meter's 60 nS range measures conductance in
nanosiemens (1 nS = 0.000000001 Siemens). Because
such small amounts of conductance correspond to
extremely high resistance, the nS range lets you
determine the resistance of components up to 100,000
M, 1/1 nS = 1,000 M.
To measure conductance, set up the Meter as shown for
measuring resistance (Figure 5); then press Cuntil
the nS indicator appears on the display.
The following are some tips for measuring conductance:
High-resistance readings are susceptible to
electrical noise. To smooth out most noisy readings,
enter the MIN MAX recording mode; then step to the
average (AVG) reading.
There is normally a residual conductance reading
with the test leads open. To ensure accurate
readings, use the relative (REL) mode to subtract
the residual value.

Other manuals for Fluke 87V Ex

Related product manuals