True-rms Multimeter
Maintenance
39
Table 8. Approved Batteries
Battery Description Manufacturer Type
Alkaline Energizer
Energizer No. 522
Eveready 6LR61
Alkaline Daimon 6LR61
Alkaline
Alkaline Ultra
Professional Alkaline Battery
Procell
Plus MN1604 6LR61
Procell MN1604 6LR61
Ultra M3 MN1604 6LR61
Duracell 6LR61
Ucar Gold 6LR61 Energizer 6LR61
Alkaline 4822
Alkaline Universal No. 4022
Alkaline Electric Power
No. 8022
Electric Power No.8022
High Energy No.4922
Industrial Alkaline No.4022
Varta 6LR61
Table 8. Approved Batteries (cont.)
Battery Description Manufacturer Type
Alkaline Power Line Industrial
Battery
Industrial Alkaline 6LR61
Powermax 6LR61
Panasonic 6LR61
Super Alkaline 1604A GP 6LR61
Replacing the Fuses
W Warning
Do not remove or install fuses while in Ex-
hazardous areas.
To avoid personal injury or damage to the
Meter, install only approved fuses shown in
Table 9. The fuses are an integral part of the
Meter’s IS protection circuits. Using non
approved fuses voids the ATEX safety
certification.
Referring to Figure 11, examine or replace the Meter's
fuses as follows:
1. Turn the rotary switch to OFF and remove the test
leads from the terminals.