True-rms Multimeter
Detailed Specifications
53
MIN MAX Recording
Nominal Response Accuracy
100 ms to 80 %
Specified accuracy ± 12 counts for changes > 200 ms in duration
(± 40 counts in ac with beeper on)
100 ms to 80 %
(dc functions)
120 ms to 80 %
(ac functions)
250 µs (peak)
[1]
Specified accuracy ± 12 counts for changes > 200 ms in duration
Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of range
Specified accuracy ± 100 counts for changes > 250 µs in duration
(add ± 100 counts for readings over 6000 counts)
(add ± 100 counts for readings in Low Pass mode)
[1] For repetitive peaks: 1 ms for single events.