Supported SCPI Commands
47
[SENSe:]FILTer[:DC]:DIGital[:STATe](?) <n>
Description Activates or deactivates the digital averaging filter to improve noise
immunity for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the digital averaging filter off.
ON or 1 Turns the digital averaging filter on.
Examples FILT:DIG ON Turns the digital averaging filter on.
Query FILT:DIG? Returns the digital averaging filter setting. (0 =
OFF and 1 = ON)
[SENSe:]VOLTage[:DC]:FILTer[:STATe](?) <n>
Description Activates or deactivates the 3-pole analog filter to improve noise immunity
for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the analog dc voltage filter off.
ON or 1 Turns the analog dc voltage filter on.
Examples VOLT:FILT ON Turns the analog dc voltage filter on.
Query VOLT:FILT? Returns the analog dc voltage filter setting.
[SENSe:]VOLTage[:DC]:FILTer:DIGital[:STATe](?) <n>
Description Activates or deactivates the digital averaging filter to improve noise
immunity for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the digital averaging filter off.
ON or 1 Turns the digital averaging filter on.
Examples VOLT:FILT:DIG ON Turns the digital averaging filter on.
Query VOLT:FILT:DIG? Returns the digital averaging filter setting. (0 =
OFF and 1 = ON)
[SENSe:]CURRent[:DC]:FILTer[:STATe](?) <n>
Description Activates or deactivates the 3-pole analog filter to improve noise immunity
for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the analog dc voltage filter off.
ON or 1 Turns the analog dc voltage filter on.
Examples CURR:FILT ON Turns the analog dc voltage filter on..
Query CURR:FILT? Returns the analog dc voltage filter setting. (0 =
OFF and 1 = ON)