8845A/8846A
Programmers Manual
48
[SENSe:]CURRent[:DC]:FILTer:DIGital[:STATe](?) <n>
Description Activates or deactivates the digital averaging filter to improve noise
immunity for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the digital averaging filter off.
ON or 1 Turns the digital averaging filter on.
Examples CURR:FILT:DIG ON Turns the digital averaging filter on.
Query CURR:FILT:DIG? Returns the digital averaging filter setting. (0 =
OFF and 1 = ON)
[SENSe:]RESistance[:DC]:FILTer[:STATe](?) <n>
Description Activates or deactivates the 3-pole analog filter to improve noise immunity
for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the analog dc 2-wire resistance filter
off.
ON or 1 Turns the analog dc 2-wire resistance filter
on.
Examples RES:FILT ON Turns the analog dc 2-wire resistance filter on.
Query RES:FILT? Returns the analog dc 2-wire resistance filter
setting. (0 = OFF and 1 = ON)
[SENSe:]RESistance[:DC]:FILTer:DIGital[:STATe](?) <n>
Description Activates or deactivates the digital averaging filter to improve noise
immunity for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the digital averaging filter off.
ON or 1 Turns the digital averaging filter on.
Examples RES:FILT:DIG ON Turns the digital averaging filter on.
Query RES:FILT:DIG? Returns the digital averaging filter setting. (0 =
OFF and 1 = ON)
[SENSe:]FRESistance[:DC]:FILTer[:STATe](?) <n>
Description Activates or deactivates the 3-pole analog filter to improve noise immunity
for dc functions. This setting is stored in volatile memory.
Parameters <n> = OFF or 0 Turns the analog dc 4-wire resistance filter
off.
ON or 1 Turns the analog dc 4-wire resistance filter
on.
Examples FRES:FILT ON Turns the analog dc 4-wire resistance filter on.
Query FRES:FILT? Returns the analog dc 4-wire resistance filter
setting. (0 = OFF and 1 = ON)