EasyManua.ls Logo

Fluke CombiScope PM3370B - Page 121

Fluke CombiScope PM3370B
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND REFERENCE 4 - 21
*
OPT? Option identification
Syntax:
*
OPT?
Response:
<option> {,<option>}
<option> <name>:<serial_nr>:<sw_level>
<name> IEEE | EXT | EM | MP
<serial_nr> Serial number is always 0.
<sw_level> Software level is always 0.
Description:
The
*
OPT? query reports which options are present.
If <option> = IEEE:0:0, the IEEE-488.2/SCPI option is installed.
If <option> = EXT:0:0, the EXTernal trigger option is installed.
If <option> = EM:0:0, Extended Memory is available.
If <option> = MP:0:0, the Math Plus option is installed.
Example:
Send
*OPT?
Read
IEEE:0:0,MP:0:0 The IEEE and MathPlus option are
available.
Front panel compliance:
The
*
OPT? query is the remote equivalent of the Maintenance option of the
UTILITY menu.

Table of Contents

Related product manuals