4 - 22 COMMAND REFERENCE
*
RCL Recall instrument setup
Syntax:
*
RCL <numeric_data>
Description:
The
*
RCL command restores instrument settings from one of the internal memory
registers 0 .. 10. The settings in memory register 0 are standard settings, which
can only be recalled. The settings in the memory registers 1 through 10 are
programmable by sending the
*
SAV command.
After power on the current settings, just before power off, are restored. These
current settings are saved in non-volatile memory (battery backed-up).
Example:
Send
*SAV 2 Stores the actual instrument settings into
memory register 2.
.
.
Send
*RCL 2 Restores the instrument settings from memory
register 2.
Front panel compliance:
The
*
SAV/
*
RCL commands are the remote equivalent of the front panel softkey
operation via the SETUPS/RECALL menu. The standard settings stored in
memory 0 can be changed via the front panel FRONT SETUPS menu.