Selecting the M Numerical Value. Press the
Mx+B key to access the Mx+B menu. Press the
up/down and left/right arrow keys to enter a five
digit number that defines the numerical value for
the measurement multiplier, ignoring the decimal
point or scale multiplier. The column being
configured has a bright display. For example, for
an M of 1000 (1k), enter +01000. When the
entry is correct, press the ENTER key.
Selecting the M Decimal Value. Press the
left/right arrow keys to position the decimal point
in the number selected in the previous step. For
the example above, the setting would be +01.000.
When the decimal point is correct, press the
ENTER key.
Selecting the M Scale Modifier. Press the
up/down arrow keys to select the desired scale
modifier: X1 (x1), m (x .001), M (x1,000,000), or k
(x1,000). For the example above, the +01.000
scale modifier would be k. When the scale
modifier is correct, press the ENTER key.
± 0 0 0 0 0
- to -
± 9 9 9 9 9
SET Mx+B
SET Mx+B
Mx+B
A
(Continued on the next page)
op26f.eps
Figure 2-14. Setting the Mx+B Scaling