Calibration Manual
Verification Tests
21
Table 7. Resistance Measurement Test Using a 5500A or 5520A Calibrator
787 Range
Calibrator
Resistance
Calibrator
Compensation Mode
Minimum
 Reading
Maximum
Reading
400 Ω 120 Ω 2-Wire 119.6 Ω 120.4 Ω
400 Ω 300 Ω 2-Wire 299.2 Ω 300.8 Ω
4 kΩ 1.2 kΩ 2-Wire 1.197 kΩ 1.203 kΩ
4 kΩ 3 kΩ 2-Wire 2.993 kΩ 3.007 kΩ
40 kΩ 12 kΩ 2-Wire 11.97 kΩ 12.03 kΩ
40 kΩ 30 kΩ 2-Wire 29.93 kΩ 30.07 kΩ
400 kΩ 120 kΩ OFF 119.7 kΩ 120.3 kΩ
400 kΩ 200 kΩ OFF 199.5 kΩ 200.5 kΩ
400 kΩ 300 kΩ OFF 299.3 kΩ 300.7 kΩ
4 MΩ 1.2 MΩ OFF 1.993 MΩ 1.207 MΩ
4 MΩ 3.0 MΩ OFF 2.986 MΩ 3.014 MΩ
40 MΩ 12 MΩ OFF 11.67 MΩ 12.33 MΩ
40 MΩ 30 MΩ OFF 29.22 MΩ 30.78 MΩ