Calibration Manual
Verification Tests
21
Table 7. Resistance Measurement Test Using a 5500A or 5520A Calibrator
787 Range
Calibrator
Resistance
Calibrator
Compensation Mode
Minimum
Reading
Maximum
Reading
400 120 2-Wire 119.6 120.4
400 300 2-Wire 299.2 300.8
4 k 1.2 k 2-Wire 1.197 k 1.203 k
4 k 3 k 2-Wire 2.993 k 3.007 k
40 k 12 k 2-Wire 11.97 k 12.03 k
40 k 30 k 2-Wire 29.93 k 30.07 k
400 k 120 k OFF 119.7 k 120.3 k
400 k 200 k OFF 199.5 k 200.5 k
400 k 300 k OFF 299.3 k 300.7 k
4 M 1.2 M OFF 1.993 M 1.207 M
4 M 3.0 M OFF 2.986 M 3.014 M
40 M 12 M OFF 11.67 M 12.33 M
40 M 30 M OFF 29.22 M 30.78 M