Calibration Manual
Verification Tests
25
Frequency Measurement Test
1. Put the calibrator in Standby (STBY) mode.
2. Put the UUT rotary switch in the
K (ac volts) position.
3. Press
F to toggle to the frequency measurement function.
4. Connect the calibrator to the
COM and z terminals on the UUT as shown in
Figure 11.
5. Apply the values from the calibrator shown in Table 11 and compare the readings on
the UUT to the acceptable readings shown.
Table 11. Frequency Measurement Test
787 Range Calibrator Voltage
and Frequency
Minimum Acceptable
Reading
Maximum Acceptable
Reading
199.99 Hz 5 V @ 100 Hz 99.98 Hz 100.02 Hz
1999.9 Hz 5 V @ 1000 Hz 999.8 Hz 1000.2 Hz
19.999 kHz 5 V @ 10 kHz 9.998 kHz 10.002 kHz
OFF
A
mA
COM
V
1000V
30mA
FUSED
0.44A
(1A/30 sec)
FUSED
mA
mA
A
mA
OUTPUT 0-24mA
SOURCE SIMULATE
+ +
% STEP COARSE FINE
mV
V
V
OUTPUT
CAT
787
PROCESSMETER
MIN MAX RANGE HOLD
H
REL Hz
UUT
5500A
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
LT010F.EPS
Figure 11. Frequency Measurement Test Connections

Other manuals for Fluke ProcessMeter 787

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ProcessMeter 787 and is the answer not in the manual?

Fluke ProcessMeter 787 Specifications

General IconGeneral
BrandFluke
ModelProcessMeter 787
CategoryMultimeter
LanguageEnglish

Related product manuals