ProcessMeter™
Required Equipment
9
Required Equipment
Equipment and software required to perform the procedures in this manual are identified
in Table 2.
If the recommended equipment model is not available, in some cases other equipment
can be substituted as long as it meets the specifications indicated.
Warning
To avoid safety hazards and equipment damage during the calibration
procedure, use the specified calibration equipment listed in Table 2.
Using unspecified equipment can jeopardize the calibration
verification test and pose safety hazards.
Note
Unless otherwise indicated, all connection diagrams for the calibration
verification tests in this manual showing a calibrator or digital multimeter
use a Fluke 5522A calibrator, Fluke 8508A Reference Multimeter, Keysight
3458A DMM, or equivalent.
If you are using a different calibrator or DMM, make the connections
appropriate for that instrument.
Table 2. Required Equipment and Software
Equipment Minimum Specifications Recommended Model
Calibration Source No Substitute Fluke Model 5522A or
equivalent
Digital Multimeter No Substitute Fluke 8508A,
Keysight 3458A
Test Leads, low leakage,
RG-58/U type
Leakage resistance > than 1.0 x 10
13
Ω at
45 °C and 75 % relative humidity
Fluke 5440A-7002 Low
Thermal Test Leads
1-kΩ shunt 1 kΩ, 1 %, 2 watts, Low TC is preferable ---