ProcessMeter™
Calibration Adjustment
31
Milliamps DC Adjustment
1. Connect the ProcessMeter to the mA output of the 5522A calibrator.
2. Turn the UUT’s switch to
W. Make sure the test leads are in the mA and COM
inputs.
3. Press and hold the Calibration Button for approximately 2 seconds. The unit will beep
(see Figure 11).
Note
Pressing the Calibration Button puts the ProcessMeter into and out of
calibration mode. The ProcessMeter will remain in calibration mode until
the unit is turned off or the calibration button is pressed a second time.
CAL appears in the bottom display when the ProcessMeter is in calibration
mode.
4. Apply 0 mA dc. Press after the sourced value appears. Do not alter the
sourced value while the display reads Busy.
5. Apply 30 mA dc. Press after the sourced value appears. Do not alter the
sourced value while the display reads Busy.
6. When Store is displayed, press to store the calibration value.
Amps DC Adjustment
1. Connect the ProcessMeter to the A output of the 5522A calibrator.
2. Turn the UUT’s switch to
W. Make sure the test leads are in the A and COM jacks.
3. Press and hold the Calibration Button for 2 seconds (see Figure 11). The unit will
beep.
4. Apply 0 A dc. Press after the reading stabilizes.
5. Apply 1 A dc. Press after the reading stabilizes.
Caution
Remove 1 A from UUT promptly after storing calibration constant.
Fuse will blow after 30 seconds.
6. Press to store calibration constants.

Other manuals for Fluke ProcessMeter 789

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ProcessMeter 789 and is the answer not in the manual?

Fluke ProcessMeter 789 Specifications

General IconGeneral
Diode testYes
Frequency range0 - 19.999 kHz
AC current range0 - 1 A
AC voltage range0 - 1000 V
Resistance range0 - 40000000 Ω
Resolution (AC voltage)0.1 mV
Dimensions (WxDxH)100 x 203 x 50 mm
Weight and Dimensions IconWeight and Dimensions
Weight610 g

Related product manuals