EasyManua.ls Logo

Fluke SimpliFiber Pro - Cleaning Connectors and Adapters; Cleaning Bulkhead Connectors; Cleaning the Meters Connector; Cleaning Fiber Adapters

Fluke SimpliFiber Pro
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning Connectors and Adapters
15
Cleaning Connectors and Adapters
Always clean and inspect fiber connectors before making
connections. Use fiber optic solvent and optical-grade wipes
or swabs to clean connectors as follows:
Cleaning Bulkhead Connectors (sources and
patch panels)
1
Touch the tip of a fiber optic solvent pen or swab
soaked in solvent to a lint-free dry wipe or fiber
cleaning card.
2
Touch a new, dry swab to the solvent spot on the wipe
or card.
3
Push the swab into the connector, twist it around 3 to 5
times against the end-face, then remove and dispose of
the swab.
4
Dry the connector with a dry swab by twisting it
around in the connector 3 to 5 times.
5
Inspect connectors with a fiber microscope, such as the
Fluke Networks FiberInspector
Video Microscope
before making connections.
Cleaning the Meter’s Connector
1
Remove the connector adapter to expose the
photodiode lens (see Figure 4).
2
Use the method described in steps 1 and 2 above to
dampen a swab with solvent.
3
Twist the damp swab around against the lens 3 to 5
times; then twist a dry swab around against the lens 3
to 5 times.
Cleaning Fiber Adapters
Periodically clean fiber adapters with a swab and fiber optic
solvent. Dry with a dry swab before use.