There are three measurement mode positions associated with this switch:
Automatic Semi Automatic Manual
In Automatic mode the Ti30 imager automatically adjusts the image to show the
lowest temperature value present in the thermal scene (MIN) and the highest temperature
value in the scene (MAX).
In
Semi-Automatic mode the Ti30 imager keeps calculating the MIN limit (lowest
temperature value on the thermal scene) automatically.
In
Manual mode the user can adjust Gain and Level manually.
See Selecting the Measurement Mode on page 58 for more information. The default position
is automatic mode.
Gain and Level Thumbwheels
The Level Thumbwheel control adjusts the median point of a given temperature
scale.
The
Gain Thumbwheel control adjusts the temperature band or range around the
median Level point.
See Gain and Level Thumbwheels on page 61 for more information about these controls.