EasyManua.ls Logo

Force 5100ex - Calibrating Axle Pulses by Distance; To Calibrate the Axle Pulses by Distance

Force 5100ex
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5100ex Calibration Manual
Calibrating Axle Pulses By Distance
29
Calibrating Axle Pulses By Distance
The 5100ex can automatically set the Calibration → Speed → Axle Pulses value by recording the
number of pulses traveled in a mile.
Before calibrating the Axle Pulses by distance:
1. Set the Calibration → Speed → Speedo Type setting.
2. Set the Calibration → Speed → Low Trip Point setting.
3. Set the Calibration → Speed → High Trip Point setting.
If the settings above are not configured properly, the axle pulses may be recorded incorrectly.
To calibrate the Axle Pulses by distance:
Calibrating axle pulses can be completed using either English or Metric units. The steps below
will walk you through calibrating the axle pulses using English units.
STEP 1: Enter the Calibration Menu. See Entering the Calibration Menu on page 2.
STEP 2: Navigate to the Calibration → Speed → Cal By Distance setting. Figure 2 will
appear, as shown below.
Figure 2: Cal By Distance Screen
STEP 3: Press the Continue button.
STEP 4: Drive the vehicle to a landmark from which you can measure exactly one mile
such as an Interstate Highway mile marker.
STEP 5: Press the green, Standby knob to begin counting axle pulses. The status should
change to “Counting”.
STEP 6: Drive exactly one mile.

Table of Contents