Forma Scientific, Inc._____________________________________________Service
6-2
Symptom Possible Cause
Problem 7: Digital CO
2
display and
FYRITE read more than 1% different from
setpoint. CO
2
alarm not on.
Alarm is disabled.
Defective alarm circuit.
Defective CO
2
sampler or control.
Problem 8: CO
2
display and FYRITE read
0% CO
2
. Setpoint is OK. CO
2
alarm is
activated. CO
2
inject light is on.
Loss of CO
2
supply.
Defective CO
2
solenoid.
Clogged CO
2
in-line filter.
Problem 9: CO
2
display shows some
random number or decimal point.
Defective readout board.
Defective output to readout board from
CO
2
control.
Faulty interconnecting wiring.
Problem 10: CO
2
display will not go to
zero no matter how long the door is left
open.
Improper procedure. Close door and allow
temperature and RH to stabilize eight
hours.
Defective CO
2
sensor.
Defective CO
2
control.
Problem 11: Display flickers badly or
counts up and down three or four digits.
Defective readout board.
Excessive electrical interference near the
cabinet.
Faulty grounding circuit.
Sampler blower is improperly installed or is
defective.