Forma Scientific, Inc._____________________________________________Service
6-1
Section 6 - Service
6.1 General Troubleshooting
Symptom Possible Cause
Problem 1: Pilot lights not on. Readouts
are dark. CO
2
power switch is on.
Unit unplugged, or main power switch not on.
Circuit breaker tripped.
Outlet overload protection has tripped.
No voltage at outlet.
Defective incubator wiring.
Problem 2: Pilot lights on but heat light
and alarm monitor module lights out.
0.3 Amp fuse in control panel is blown.
Problem 3: CO
2
system indicator and
readout are dark no control of CO
2
in
chamber. Switch is on.
Circuit breaker tripped.
Defective switch or wiring.
Defective CO
2
module.
Problem 4: CO
2
setpoint and display
agree. FYRITE reads lower.
FYRITE not zeroed.
Wool filter in FYRITE is dry.
FYRITE fluid need to be changed.
Chamber absolute humidity has decreased.
Control is incorrectly zeroed.
Defective CO
2
sensor or control.
Problem 5: Alarm sounds periodically,
even with the alarm disabled. CO
2
alarm
light not activated.
Over temp control set too low.
Shorted heater.
Defective temp control or sensor.
Problem 6: CO
2
alarm and light are on.
Alarm can be reset or disabled. If reset,
alarm activates again.
CO
2
level has deviated more than 1% from
setpoint.
CO
2
supply has been interrupted.
CO
2
inject circuit is defective.
CO
2
solenoid is defective.
CO
2
sensor fan is defective.