Forma Scientific, Inc.______________________________Specifications/Accessories
7-2
7.3 Temperature Control
• Chamber temperature is constantly displayed on easy-to-read LCD.
• Independent, electronic, dual temperature controller maintains chamber
temperature at a fixed +37
o
C or adjustable from +5
o
C above ambient to +60
o
C.
• CH/P module maintains overtemperature safety limit independent of temperature
controller for true safety back-up. Limits are adjustable in 0.1
o
C increments.
Audible and visual alarms warn of temperature deviations. Audible alarm can be
silenced with 30 minute ringback.
• Tamper-resistant controls prevent alteration of overtemperature safety settings.