Forma Scientific, Inc._____________________________________________Service
6-10
6.7 Replacing the Temperature Control
(Refer to Wiring Diagram, 140179-71-0-D)
Turn the incubator off and disconnect it from the power source.
Remove the screws securing the right side access panel. Remove the panel
and set aside.
Remove the temperature control knob by loosening the two Allen head
screws recessed in the knob.
Remove the nut on the shaft on the outside of the panel, then remove the nut
on the back of the panel behind the control knob.
Remove the wiring to the temperature control and note the connections.
Remove the control and replace the new unit by reversing the above
procedure.
6.8 Replacing Pilot Lights
Turn the incubator off and disconnect it from the power source.
Remove the screws securing the right side access panel. Remove the panel
and set aside.
Disconnect the wiring behind the defective light and break the retaining clip
loose. Remove the pilot light.
Insert the new pilot light from the front of the panel and install the new
retaining clip.
Reconnect the wiring.