98
Factory default descriptions
The channels have the following factory default descriptions:
Combinining pressures
All invasive pressure waveforms can be displayed together so that 
they use an area of two normal waveforms, or so that all are 
combined in the same field with the same zero line.
1. Press the Monitor Setup key.
2. Select Screen Setup.
3. Select Waveform Fields.
• To combine all pressure waveforms in one field, select 
Combine Pressures to YES.
Determining pressure values visually
By moving the horizontal cursor across the pressure waveform, you 
can get accurate pressure values at selected points. This may be 
useful, for example, if the patient's breathing pattern is irregular. The 
cursor is not available for pressures shown with a combined scale.
1. Press the IBP key.
2. Select IBP1 Setup - IBP1 Cursor and IBP2 Setup - IBP2 Cursor
3. Move the cursor up or down by turning the Trim Knob. Every time 
the cursor is moved, the time (hours and minutes) and pressure 
values appear on the screen. In this way you can keep track of 
the changes made.
4. You can remove the cursor by selecting Remove Cursor. Note 
that if the cursor is not removed, it remains visible on the Normal 
Screen.
LABEL IBP1, Art, ABP IBP2, CVP RAP, LAP ICP PA RVP UAC UVC
Scale 
mmHg/kPa 200/30 20/4 20/4 20/4 60/8 60/8 100/14 10/2
Color Red Blue White White Yellow White Red White
Alarm source Sys Mean Off Off Off Off Sys, Dia, 
Mean
Mean
Digit format S/D Mean Mean CPP S/D S/D S/D Mean
Filter (Hz) 22999991414
Response Normal Normal Normal Normal Normal Normal Normal Normal