124
Messages
Always check the patient first. If any problem or message persists, contact qualified service personnel.
Messages are listed here in alphabetical order.
•Agent Mixture
− Mixture of halogenated agents is detected. Check the
ventilator and agent vaporizer settings.
•Alarm setup changed from Central
− Check the alarm limits and the arrhythmia alarm priorities,
see “Alarms” and “ECG.”
• Alarms silenced from Central
− If required, turn on the bedside alarms through Alarms
Setup
- Audio ON/OFF - Activate Alarms.
• Alarm volume changed
− The network connection is lost, and the local alarm volume is
increased. Readjust volume if desired.
• Apnea/No Breath
− Check the patient status.
− Check the ventilator and breathing circuit.
• Arrhythmia Paused
− ECG: Check the patient status. Check the electrodes.
• Art disconnect/ABP disconnect/UAC disconnect
− IBP: Check the patient status.
− Check the Art/ABP/UAC line.
• Asystole
− Check the patient status.
− Check the electrodes.
• Battery A temperature high, Battery B temperature high
− Replace the battery.
− If the problem persists, contact authorized service personnel.
• Battery empty, Battery low
− Replace the battery, see “Replacing the batteries”, or connect
the monitor to power outlet.
•Brady
− Check the patient status.
• Call service: UMBC error
− UMBC: Contact authorized service personnel.
• Check D-Fend
− Check that the water trap is properly attached to the module.
•Check OCRG
− New OCRG snapshot is created, please check.
•Check NIBP
− Check the patient status.
− Check the measurement setup.
− Check the cuff and hoses.