EasyManua.ls Logo

GE B20 - Page 129

GE B20
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
125
•Check SpO2 probe
SpO
2
: Check the sensor and connections.
Check sample gas out
Gases: Remove blockage from the sample gas outlet.
Condition Battery A, Condition Battery B
Condition the battery according to the instructions of the
external charger.
Creating OCRG snapshot
In neonatal mode, the value of HR, SpO
2
or Apnea time is out
of limits.
Default settings returned.
Check your power and reset the settings.
•DEMO MODE
DEMO mode is active.
ECG module error
Contact authorized service personnel.
End of 20 min trend data
Check the graphical trends and move the cursor.
•Entropy cable off
Connect the Entropy cable to the Entropy module.
Entropy sensor off
Check that the sensor is properly attached to the patient.
Entropy sensor check failed
Check sensor placement and attachment.
Press each electrode in the sensor.
Replace the sensor.
Frame temperature high
Make sure there is sufficient ventilation.
If the problem persists, contact authorized service personnel.
Turn off the monitor, wait for it cool down.
•Freq PVCs
Check the patient status.
Incompatible SpO2 Probe
Replace the sensor.
If the problem persists, contact authorized service personnel.
InvBP's not zeroed
Zero the channel indicated or zero both channels.
•Leads off
ECG: Reconnect the disconnected trunk cable, electrode or
leadwire. Change the trunk cable, lead set and module.
License invalid
Contact authorized service personnel to set up the monitor’s
right license again.
Loading from network
Loading mode from network.
Loading failed
Load mode from network failed.

Related product manuals