125
•Check SpO2 probe
− SpO
2
: Check the sensor and connections.
• Check sample gas out
− Gases: Remove blockage from the sample gas outlet.
• Condition Battery A, Condition Battery B
− Condition the battery according to the instructions of the
external charger.
• Creating OCRG snapshot
− In neonatal mode, the value of HR, SpO
2
or Apnea time is out
of limits.
• Default settings returned.
− Check your power and reset the settings.
•DEMO MODE
− DEMO mode is active.
• ECG module error
− Contact authorized service personnel.
• End of 20 min trend data
− Check the graphical trends and move the cursor.
•Entropy cable off
− Connect the Entropy cable to the Entropy module.
• Entropy sensor off
− Check that the sensor is properly attached to the patient.
• Entropy sensor check failed
− Check sensor placement and attachment.
− Press each electrode in the sensor.
− Replace the sensor.
• Frame temperature high
− Make sure there is sufficient ventilation.
− If the problem persists, contact authorized service personnel.
− Turn off the monitor, wait for it cool down.
•Freq PVCs
− Check the patient status.
• Incompatible SpO2 Probe
− Replace the sensor.
− If the problem persists, contact authorized service personnel.
• InvBP's not zeroed
− Zero the channel indicated or zero both channels.
•Leads off
− ECG: Reconnect the disconnected trunk cable, electrode or
leadwire. Change the trunk cable, lead set and module.
• License invalid
− Contact authorized service personnel to set up the monitor’s
right license again.
• Loading from network
− Loading mode from network.
• Loading failed
− Load mode from network failed.