126
• Long measurement time
− NIBP: reduce patient’s motion.
• Low gas sample flow
− Check the patient status.
− Change the sampling line or water trap.
•Mark x
− A snapshot is created manually.
•Memory ERROR
− Contact authorized service personnel.
•Mode Data Reset
− Contact authorized service personnel.
• Module power supply overload
− Check the patient status.
− Contact authorized service personnel.
• Multiple agents present
− Check if the anesthetic agents are more than one.
− Check the patient status.
•Network: x
− Network is connected.
• Network down: xxx (xxx = network name)
− Try to re-establish the connection.
− Contact authorized service personnel.
• Network recording...
− Network recorder has been started. Please wait until the 
recording is finished.
• NIBP call service error
− NIBP: Contact authorized service personnel.
•NIBP manual
− Check the cuff and cuff hose.
•NIBP cuff overpressure
− Check NIBP cuff hose and tubes.
− Restart measurement.
• NIBP cuff loose
− Check the cuff and cuff hose.
• NIBP cuff occlusion
− Check the tubes and hose.
•No battery backup
− Replace the batteries, see “Replacing the batteries.”
• No Entropy sensor
− Check connection between Entropy sensor and cable.
•No Transducer
− IBP: Connect the transducer or the cable.
•No SpO2 probe
− Check connection between the SpO
2
 sensor and module.
•No SpO2 pulse
− Try other measuring sites.