126
Long measurement time
NIBP: reduce patient’s motion.
Low gas sample flow
Check the patient status.
Change the sampling line or water trap.
•Mark x
A snapshot is created manually.
•Memory ERROR
Contact authorized service personnel.
•Mode Data Reset
Contact authorized service personnel.
Module power supply overload
Check the patient status.
Contact authorized service personnel.
Multiple agents present
Check if the anesthetic agents are more than one.
Check the patient status.
•Network: x
Network is connected.
Network down: xxx (xxx = network name)
Try to re-establish the connection.
Contact authorized service personnel.
Network recording...
Network recorder has been started. Please wait until the
recording is finished.
NIBP call service error
NIBP: Contact authorized service personnel.
•NIBP manual
Check the cuff and cuff hose.
•NIBP cuff overpressure
Check NIBP cuff hose and tubes.
Restart measurement.
NIBP cuff loose
Check the cuff and cuff hose.
NIBP cuff occlusion
Check the tubes and hose.
•No battery backup
Replace the batteries, see “Replacing the batteries.”
No Entropy sensor
Check connection between Entropy sensor and cable.
•No Transducer
IBP: Connect the transducer or the cable.
•No SpO2 probe
Check connection between the SpO
2
sensor and module.
•No SpO2 pulse
Try other measuring sites.