General specifications
Size
B40 and B20 monitor
Without extension
module
312±5 mm (H) * 312±5 mm (W) * 158±5
mm (D)
With extension module 312±5 mm (H) * 352±5 mm (W) * 178±5
mm (D)
Weight
B40 with extension
module
7 kg
B20 with extension
module
6 kg
Monitor Environment
Operating temperature Normal operation: 5 to 40°C (41 to
104°F)
Charging batteries: 5 to 35°C (41 to
95°F)
Storage and transport
temperature
-20 to 60°C (-4 to 140°F)
Operating humidity 20 to 90% noncondensing
Storage and transport
humidity
10 to 90% noncondensing
Operating atmospheric
pressure
700 to 1060 hPa
(525 to 795 mmHg)
Storage and transport
atmospheric pressure
700 to 1060 hPa
(525 to 795 mmHg)
E-miniC module Environment
Operating temperature 10 to 40°C (50 to 104°F)
Non-operating
temperature
-25 to 70°C (-13 to 158°F)
Operating humidity 10 to 95% noncondensing
Non-operating humidity 10 to 95% noncondensing
Operating altitude 666 to 1060 mbar
E-sCAiO, E-sCO, N-CAiO modules Environment
Operating temperature 10 to 40°C (50 to 104°F)
Non-operating
temperature
-25 to 60°C (-13 to 140°F)
Operating humidity 10 to 98% noncondensing
Non-operating humidity 10 to 90% noncondensing