127
• No printer selected
− Select a printer, see the “Printing and recording“ chapter.
• Patient discharged
− Discharge patient.
• Patient admitted
− Admit patient.
• P1 over range/P2 over range
− Check the patient status.
• P1 under range/P2 under range
− Check the patient status.
− Check the cables.
− Rezero the transducer.
− Change the module.
− If the problem persists, contact authorized service personnel.
• Printer error
− Printer is not working properly. Check that the network
printer is operational.
• Printing...
− Printing on network printer has been started. Please wait
until the printing is finished.
• Printing ready
− Monitor has completed sending printing data to the printer.
Please wait until the printing is finished.
• Recorder: out of paper
− Add paper.
• Recorder: cover open
− Close the recorder cover.
• Recorder: input voltage high, Recorder: input voltage low
− Contact authorized service personnel.
• Recorder: thermal array overheat
− Contact authorized service personnel.
• Recorder module removed
− Reconnect the recorder module if you need a recorder.
• Recorder: system error 29
− The message rarely appear when the record information too
large, the current recording will stop also. Please press the
Recorder Start/Stop key again.
• Recorder: system error X
− Contact authorized service personnel.
• Replace Battery A, Replace Battery B
− Replace the battery as soon as possible, see “Replacing the
batteries.”
•Replace D-Fend
− Gases: Replace the water trap.
• Restart needed
− The monitor should be restarted.
• Saving to network
− Saving trend and information to Network
• Sample line blocked
− Gases: Change the sampling line and water trap.