128
• Select inflation limits
− NIBP: You are using a hose without an automatic
identification. Select appropriate inflation limits.
NOTE: AUTO option is not available for these hoses.
• Snapshot created
− A snapshot is created manually or on alarms.
• Snapshot memory full. Oldest snapshot erased.
− The oldest snapshot lost.
• SpO2 probe off
− Check connection between sensor and patient.
− Replace the sensor.
•SpO2 faulty probe
− Replace the sensor.
•SpO2 module error
− Contact authorized service personnel.
•Tachy
− Check the patient status.
•Temp sensor error
− Contact authorized service personnel.
• T1 temperature error/T2 temperature error
− Check the probe whether is set to 400.
− Contact authorized service personnel.
• V Fib
− Check the patient status.
•V Tach
− Check the patient status.
• Weak pulsation
− Check the patient status.
− Check the NIBP cuff position and attachment.
− Check that the cuff is not damaged.
− Repeat the measurement and check the patient status.
• Zero ICP separately
− In IBP menu to zero ICP labelled channel separately.
• xxx high/low (xxx = measurement parameter)
− Check the patient status.
− Adjust the alarm limits.
• xxx measurement removed
− Acquisition module has been removed. Connect the module
if you want to restart the measurement.
− Check the patient status.